Advancing Anti-HCV Point-of-Care Testing: Fourteen Years of Quality Insights from Brazil’s EQAP (2010–2024)

Advancing Anti-HCV Point-of-Care Testing: Fourteen Years of Quality Insights from Brazil’s EQAP (2010–2024)

Background - Aim

Hepatitis C virus (HCV) infection remains a global health challenge, causing chronic liver disease and approximately 400,000 deaths annually, with a high burden in low- and middle-income countries (LMICs). Anti-HCV detection is a standard screening tool, and point-of-care tests (POCT) offer rapid results to support timely decisions. External Quality Assessment Programs (EQAP) are vital for ensuring the reliability and comparability of test results.

Methods

Data from 2010 to 2024, spanning four annual rounds, were analyzed. Participants evaluated four lyophilized and/or liquid quality control samples per round. Metrics included participation, adequacy rates (%A), inadequacy rates (%I), specificity (SP), and sensitivity (SE).

Results

A total of 290 participants contributed 16,828 datasets, with participation growing from 20 laboratories in 2010 to 121 in 2024 (p<0.0000). Performance remained high over time, with overall %A 98.6%, %I 1.8%, SP 98.7%, and SE 99% (Figure 1).

Of 24 manufacturer kits, 18 achieved %A > 98%, while 2 scored below 95%. Performance highlights for commonly used kits include:

  • Manufacturer A: %A 98.3%, SP 97.9%, SE 98.9% (144 participants – 4,522 datasets)
  • Manufacturer B: %A 99.7%, SP 99.9%, SE 99.4% (94 participants – 3,133 datasets)
  • Manufacturer C: %A 99.5%, SP 99.7%, SE 99.2% (54 participants – 1,512 datasets)

Smaller datasets showed lower performance:

  • Manufacturer D: %A 94.1%, SP 92.3%, SE 97.2% (8 participants – 101 datasets)
  • Manufacturer E: %A 88.8%, SP 83.7%, SE 94.6% (6 participants – 80 datasets) (Table 1)
Participação e desempenho ao longo do tempo
Figure1: Participation and performance overtime.
Desempenho dos kits mais usados - Tabela 1
Table 1: Performance for more commonly used kits.

Conclusions

The growing adoption of POCT for anti-HCV in Brazil is evidenced by increased EQAP participation. High %A, SP, and SE, alongside low %I, confirm the reliability of POCT, matching traditional laboratory test standards. EQAP monitoring ensures robust performance and supports accurate HCV diagnostics.

References

Updated recommendations on simplified service delivery and disagnostics for hepatitis C infection: policy brief. ISBN978-92-4-005269-7(electronicversion). ISBN978-92-4-005270-3 (printversion) ©WorldHealthOrganization 2022.

C.Vassalle, A.Mercuri, S.Masini, S.Antongiovanni, A.Pilo, F.Cantini, G.C.Zucchelli. External quality assurance program in anti-HCV serological detection: an Italian experience. Immuno-analyse & Biologie Spécialisée, Volume19, Issue4, 2004, Pages 235-238, ISSN 0923-2532, https://doi.org/10.1016/j.immbio.2004.05.001.

Carvalho-Gomes Â, Cubells A, Pallarés C, Hontangas V, CondeI, Di Maira T, Peiró S, Sanfélix-Gimeno G, López-Labrador FX, Berenguer M. Apopulation-based screening for hepatitis C antibodies and active infection using a point-of-care testin a low prevalence area. PLoS One. 2020 Feb11;15(2): e0228351. doi:10.1371/journal.pone.0228351. PMID:32045417; PMCID: PMC7012430.

Authors:

A. Ferreira4, L. Bottino1, M. Souza1, L. Pereira1, E. Santiago1, A. Vieira1, L. Vasconcellos2, Á. Pulchinelli3, R. Sykora4, M. Moreira1, J. Poloni1, V. Biasoli1

1 – Controllab, Rio de Janeiro – Brasil.

2 – Faculdade de Medicina – Universidade Federal de Minas Gerais, Belo Horizonte – Brasil; Sociedade Brasileira de Patologia

Clínica/Medicina Laboratorial (SBPC/ML), Rio de Janeiro – Brasil.

3 – Grupo Fleury, São Paulo – Brasil; Sociedade Brasileira de Patologia Clínica/Medicina Laboratorial (SBPC/ML), Rio de Janeiro

See also

Sign up for our newsletter

Receive relevant publications about lab and solutions for
quality management.

plugins premium WordPress
X
We use cookies and other similar technologies to improve your experience, enabling a more personalized navigation, according to the characteristics and your interest. For more information see our Privacy and Security Policy.

PRM 0009

Certified Reference Material Producer

Accredited by Coordenação-Geral de Acreditação do Inmetro

Controllab has been accredited as a Certified Reference Material (CRM) Producer since 2016 by Cgcre, under number PMR 0009. The conformity assessment is carried out in accordance with the requirements established in ABNT NBR ISO 17034 and Inmetro standards, ensuring competence for the production of CRMs, as defined in the Scope of Accreditation.

This accreditation attests to Controllab’s technical competence in producing CRMs according to international requirements and enables it to be recognized in other countries, under mutual recognition agreements, thereby reducing technical barriers between Brazil and other countries across continents.

Materiais de Referência Certificados - PMR 0009 - ABNT NBR ISO 17034-Controllab
Cgcre is signatory to the Mutual Recognition Arrangements of the International Laboratory Accreditation Cooperation (ILAC) and Interamerican Accreditation Cooperation (IAAC).

CAL 0214

Calibration Laboratory

Accredited by Coordenação-Geral de Acreditação do Inmetro

In December 2002, the Calibration Laboratory was accredited by Cgcre. With this recognition, it became part of the Brazilian network of accredited laboratories, under number 214. Since then, it has been periodically evaluated for the maintenance and/or extension of its accreditation.

The Inmetro seal attests to Controllab’s technical competence, credibility, and operational capability for calibration services, in accordance with the requirements of ABNT NBR ISO/IEC 17025.

Calibração de Instrumentos - CAL 0214 - ABNT NBR ISO/IEC 17025 - Controllab
Cgcre is signatory to the Mutual Recognition Arrangements of the International Laboratory Accreditation Cooperation (ILAC) and Interamerican Accreditation Cooperation (IAAC).

PEP 0003

Proficiency Test Provider

Accredited by Coordenação-Geral de Acreditação do Inmetro

In September 2011, Controllab was accredited by Cgcre as a proficiency test provider, under nº PEP0003. Initially as part of a pilot project by Inmetro, this accreditation was based on ILAC G13: 2007 and included several clinical and hemotherapy tests. In the periodic evaluations of this accreditation, new segments and tests are included and the evaluation is now conducted according to ABNT NBR ISO/IEC 17043.

This accreditation attests to Controllab’s technical competence to develop and conduct proficiency tests in accordance with international requirements and enables it to be recognized in other countries, in accordance with mutual recognition agreements signed by Brazil in Mercosur, the Americas and Europe.

PEP 0003 - ABNT NBR ISO-IEC 17043 - Controllab

Cgcre is signatory to the Mutual Recognition Arrangements of the International Laboratory Accreditation Cooperation (ILAC) and Interamerican Accreditation Cooperation (IAAC).