Laboratory Performance in Molecular Detection of High-Risk HPV: Commercial Kits Outperform Laboratory-Developed Tests

Laboratory Performance in Molecular Detection of High-Risk HPV: Commercial Kits Outperform Laboratory-Developed Tests

Background - Aim

High-risk HPV causes cervical cancer, a global health challenge. Molecular DNA testing enables early detection, allowing timely treatment, improving screening, stratifying risks, and optimizing follow-ups. Reliable results require standardized laboratory performance. External Quality Assessment Programs (EQAP) ensure accuracy (AC) and harmonization in HPV diagnostics. This study aimed to evaluate the performance of laboratories in an EQAP for high-risk HPV detection, organized by a provider accredited to ABNT NBR ISO/IEC 17043:2011.

Methods

The EQAP consisted of liquid samples from cervical brushing or cell lysate. Participating laboratories received two samples quarterly to evaluate the accuracy (AC), sensitivity (SE), and specificity (SP), which were thoroughly described. Accuracy trends over time were analyzed using the Mann-Kendall test. Sub-group comparisons included in-vitro diagnostic (IVD) kits versus laboratory-developed tests (LDTs), WHO-recommended kits versus others (non-WHO), and qPCR versus non-qPCR methods. Chi-square tests and odds ratios were employed for statistical comparisons (Table 1).

Desenho do Estudo
Table 1: Study Design

Results

Between 2020 and 2024, 69 labs participated in multiple rounds of the program, with a median of 18 laboratories/round (range: 5–51), generating a total of 692 datasets (Table 2). The laboratories achieved 97.98% (678/692) AC, 97.79% (620/634) SE, and 100% (58/58) SP (Table 3). No significant trends in AC were observed over time (τ=-0.046, p=0.795). IVD kits outperformed LDTs, achieving 98.6% (612/621) AC compared to 93.0% (66/71) (OR=5.15, 95% CI 1.68–15.83, p=0.009). WHO-recommended kits and non-WHO kits demonstrated comparable AC rates of 98.4% (185/188) and 98.6% (427/433), respectively (OR=0.87, 95% CI 0.21–3.50, p=1.0). Similarly, qPCR methods showed similar performance to non-qPCR methods, with AC rates of 98.4% (541/550) versus 96.5% (136/141) (OR=2.21, 95% CI 0.73–6.70, p=0.176) (Table 4).

Visão geral do participante - Tabela 2
Table 2: Participant Overview
Desempenho ao longo do tempo - Tabela 3
Table 3: Performance over time
Comparações de subgrupos - Tabela 4
Table 4: Subgroup comparisons

Conclusions

The High-risk HPV EQAP demonstrated robust diagnostic performance, with IVD kits significantly outperforming LDTs. Both WHO-recommended and non-WHO kits, as well as qPCR and non-qPCR methods, exhibited comparable AC. These findings highlight the quality of HPV diagnostics in participating laboratories and support preferring IVD tests to address the global burden of HPV-related diseases.

References

WHO guideline for screening and treatment of cervical pre-cancer lesions for cervical cancer prevention, second edition: Use of mRNA tests for human papillomavirus (HPV)-World Health Organization, 2021.

Human papillomavirus (HPV) nucleic acid amplification tests (NAATs) to screen for cervical pre-cancer lesions and prevent cervical cancer: policy brief-World Health Organization, 2022.

Okada PA, Mitrat S, Rojanawiwat A. External quality assessment program for human papillomaviruses DNA testing in Thailand. Pract Lab Med. 2023 Dec26; 38: e00352. doi: 10.1016/j.plabm. 2023. e00352. PMID:38292923; PMCID: PMC10825476.

Authors:

G. Barra4, A. Vieira1, L. Vasconcellos2, Á. Pulchinelli3, S. Lima1, E. Santiago1, F. De Aguiar1, S. Lira1, J. Poloni1, V. Biasoli1

1 – Controllab, Rio de Janeiro – Brasil.

2 – Faculdade de Medicina – Universidade Federal de Minas Gerais, Belo Horizonte – Brasil; Sociedade Brasileira de Patologia Clínica/Medicina Laboratorial (SBPC/ML), Rio de Janeiro – Brasil.

3 – Grupo Fleury, São Paulo – Brasil; Sociedade Brasileira de Patologia Clínica/Medicina Laboratorial (SBPC/ML), Rio de Janeiro – Brasil.

4 – Sabin diagnóstico e saúde, Brasília, Brasil.

See also

Sign up for our newsletter

Receive relevant publications about lab and solutions for
quality management.

plugins premium WordPress
X
We use cookies and other similar technologies to improve your experience, enabling a more personalized navigation, according to the characteristics and your interest. For more information see our Privacy and Security Policy.

PRM 0009

Certified Reference Material Producer

Accredited by Coordenação-Geral de Acreditação do Inmetro

Controllab has been accredited as a Certified Reference Material (CRM) Producer since 2016 by Cgcre, under number PMR 0009. The conformity assessment is carried out in accordance with the requirements established in ABNT NBR ISO 17034 and Inmetro standards, ensuring competence for the production of CRMs, as defined in the Scope of Accreditation.

This accreditation attests to Controllab’s technical competence in producing CRMs according to international requirements and enables it to be recognized in other countries, under mutual recognition agreements, thereby reducing technical barriers between Brazil and other countries across continents.

Materiais de Referência Certificados - PMR 0009 - ABNT NBR ISO 17034-Controllab
Cgcre is signatory to the Mutual Recognition Arrangements of the International Laboratory Accreditation Cooperation (ILAC) and Interamerican Accreditation Cooperation (IAAC).

CAL 0214

Calibration Laboratory

Accredited by Coordenação-Geral de Acreditação do Inmetro

In December 2002, the Calibration Laboratory was accredited by Cgcre. With this recognition, it became part of the Brazilian network of accredited laboratories, under number 214. Since then, it has been periodically evaluated for the maintenance and/or extension of its accreditation.

The Inmetro seal attests to Controllab’s technical competence, credibility, and operational capability for calibration services, in accordance with the requirements of ABNT NBR ISO/IEC 17025.

Calibração de Instrumentos - CAL 0214 - ABNT NBR ISO/IEC 17025 - Controllab
Cgcre is signatory to the Mutual Recognition Arrangements of the International Laboratory Accreditation Cooperation (ILAC) and Interamerican Accreditation Cooperation (IAAC).

PEP 0003

Proficiency Test Provider

Accredited by Coordenação-Geral de Acreditação do Inmetro

In September 2011, Controllab was accredited by Cgcre as a proficiency test provider, under nº PEP0003. Initially as part of a pilot project by Inmetro, this accreditation was based on ILAC G13: 2007 and included several clinical and hemotherapy tests. In the periodic evaluations of this accreditation, new segments and tests are included and the evaluation is now conducted according to ABNT NBR ISO/IEC 17043.

This accreditation attests to Controllab’s technical competence to develop and conduct proficiency tests in accordance with international requirements and enables it to be recognized in other countries, in accordance with mutual recognition agreements signed by Brazil in Mercosur, the Americas and Europe.

PEP 0003 - ABNT NBR ISO-IEC 17043 - Controllab

Cgcre is signatory to the Mutual Recognition Arrangements of the International Laboratory Accreditation Cooperation (ILAC) and Interamerican Accreditation Cooperation (IAAC).